What Are the Different Types of K-Wires

K-wires are used for fixation of bone fragments of epi-/metaphyseal fractures, fractures of small bones, but also for reduction of fragments in multifragmental fractures to achieve stable fixation.

History of K-wire

In 1909, Martin Kirschner introduced the Kirschner wire (K-wire), initially used for skeletal traction but now serving various purposes. Its development was influenced by changing surgical goals and the introduction of antibiotics. Early K-wires were hammered into predrilled holes, but drilling became the standard insertion method due to its simplicity and benefits like percutaneous insertion. However, this technique can cause temperature elevation, leading to potential osteonecrosis and heat-related complications.

Different types of K-wire tips

Diamond Trocar Medin K wires

Drilling Efficiency and Temperature of Three Types of Kirschner Wires

Using the same drill speed (rev/min) and feed rate, the medin K-wire point produced the lowest thrust force and torque. Drill point temperatures were highest with the trochar geometry (129 ± 6°C), followed by the diamond (98 ± 7°C). The lowest temperatures were recorded with the Medin K-wire (66 ± 2°C). On repeated drilling it could be used for up to 30 holes before reaching the less satisfactory drill performance of the diamond tip. The medin K-wire provides a better alternative as it requires less effort for insertion, generates less heat and may be re-used.

Drilling Efficiency and Temperature of Three Types of Kirschner Wires 1

Graph 1 showing the thrust force and graph 2 the torque moment during the tool life test of the Medin wires in pig cortical bone (drill speed n = 280rev/min, feed rate f = 0.1mm/rev). For comparison the mean thrust force and torque moment of the diamond tip are included.

Drilling Efficiency and Temperature of Three Types of Kirschner Wires 3

Graph 3 showing the temperature elevation for different K-wire tips when drilling into pig cortical bone over 35 seconds (drill speed n = 280 rev/min, feed rate f = 0.1 mm/rev).

About CareFix

CareFix is a professional national high-tech enterprise engaged in orthopedic medical equipment design, production, sales and service. Founded in 2009, we’ve devoted to pediatric implants,foot & ankle implant,external fixation. Welcome to visit to our factory,it’s about half an hour from Shanghai Hongqiao Airport.